Infomation

Particle filter mask

Treatise 04 ธันวาคม 2563

Category :

Medical equipment / Health equipment

Product Name :

Lucca Care

Product Code :

300000013308

Category :

Manufacturer

Supplier Name :

Lucca Light

Status :

Compulsory product recall

Product type :

Foreign country

Remark/Danger :

The particle/filter retention of the material is insufficient (measured value 57,2%). Consequently, an excessive amount of particles or microorganisms might pass through the mask, increasing the risk of infection if not combined with additional protective measures.

Product Location :

Germany

Dangers found :

The particle/filter retention of the material is insufficient (measured value 57,2%). Consequently, an excessive amount of particles or microorganisms might pass through the mask, increasing the risk of infection if not combined with additional protective measures. / The product does not comply with the Personal Protective Equipment Regulation and with the relevant European standard EN 149.